sa2ei garos
CALIBRATION OF FORCE SENSORS
AND TEST PIECES

GAROS calibrate your force sensors and pieces instrumented.
We have ISO 9001 certicate and COFRAC compliance

Our capabilities of calibration
presse d'étalonnage
  • Traction / compression press
    ROELL + KORTHAUS 100 KN

  • Traction / compression press
    TESTWELL 10 KN

  • Compression press
    ENERPAC 200 KN

  • Compression press
    ENERPAC 4000 KN

 

Type
Range
Error of
accuracy between
20 % and 100 %
range
(% of point)
Repetability
error between
20 % and 100 %
range
(% of point)
Compliance
COFRAC

NF EN ISO 7500–1
Traction
10 kN
0.01
0.03
YES
Traction
50 kN
0.39
0.04
YES
Traction
100 kN
0.08
0.06
YES
Compression
5 kN
0.05
0.02
YES
Compression
10 kN
0.03
0.02
YES
Compression
20 kN
0.15
0.04
YES
Compression
50 kN
0.32
0.08
YES
Compression
100 kN
0.15
0.03
YES
Compression
200 kN
0.18
0.02
YES
Compression
500 kN
0.05
0.03
YES
Compression
4000 kN
0.01
0.02
YES

 

 

 

 

 

Sensors calibrated

  • Load cells 1 or 2 way(s)
  • "Pin" load cells 1 or 2
  • Ring Load cells
  • Multicomponents sensors
  • Pieces instrumented

Signal measured

  • mV/V : 4 ways simultaneous
  • 0-10 V : 1 way
  • 0-20 mA or 4-20 mA : 1 way
  • I.R./m in quater bridge, half bridge and wheatstone bridge : 1 way
Your specifications :

Contact us